下載
下載

下載手冊、產品規格表、軟體等等:

下載類型
機型或關鍵字

When I connect a TekConnect probe+ why doesn’t the scaling (V/D setting) on the mainframe change by the same factor as the probe?

問題:

When I connect a TekConnect probe+ why doesn’t the scaling (V/D setting) on the mainframe change by the same factor as the probe?

答案:

When the probe with the TekConnect interface is attached to the 80A03, the CSA/TDS8000B sets the V/D setting to match the dynamic range of the probe. As an example, the P7240 has 5X attenuation and +2V dynamic range. With a total dynamic range of 4V, the maximum V/D setting is 400mV/div (4V divided by 10 divisions). As a result, the V/D setting will change by a factor of 4, instead of 5.

此常見問答集適用於:

沒有任何產品系列

產品:

常見問答集 ID 60141

檢視所有常見問答集 »