What are the differences between the P5200 and the P5200A?

問題:

What are the differences between the P5200 and the P5200A?

答案:

The P5200 bandwidth is 25 Mhz, Input Impedance at the Probe Tip is 4 MΩ / 7 pF.

The P5200A bandwidth is 50 Mhz, Input Impedance at the Probe Tip is 5 MΩ / 4 pF

 

Here is an example of a fast rise time signal being measured on both probes and as you can see the new probe reduces the ringing. 

P520020to20P5200A20Compare