Does the 80A03 require any internal calibration?

問題:

Does the 80A03 require any internal calibration?

答案:

No, the 80A03 requires no internal calibration. However, after installing any electrical sampling module, the compensation procedure should be run. The compensation function is located in the Utility menu on the oscilloscope.