下載
下載

下載手冊、產品規格表、軟體等等:

下載類型
機型或關鍵字

Can the digital output pulse be longer in duration than the channel time with ExceLINX-1A?

問題:

Can the digital output pulse be longer in duration than the channel time with ExceLINX-1A?

答案:

Can the digital output pulse be longer in duration than the channel time?

Yes, but the pulse will never be reset. It will always be HI or LO depending on the configuration. It is best to select an output pulse duration that is shorter in duration than the channel time. For example if the output pulse duration is set for 100msec. Then the channel time must be set for longer than 100msec.


此常見問答集適用於:

沒有任何產品系列

產品:

常見問答集 ID 70086

檢視所有常見問答集 »