問題:
What is the difference between Position and Offset on an oscilloscope?
答案:
To answer this first you need to understand the layout of how your waveform trace is displayed.
On your oscilloscope you generally will see either 8 or 10 divisions vertically on the screen.
Along the left side of the screen you will see a marker which indicates your channels 0V level.
You can move the position of the channels 0 Voltage indicator vertically anywhere on the display.
If the signal you are probing has a voltage offset; you can compensate for the offset by applying a voltage in the scopes offset menu.